TRI-CARB 4810TR 110 V Liquid Scintillation Counter by PerkinElmer, Inc.

Manufacturer PerkinElmer, Inc.   |   Model: A481000
TRI-CARB 4810TR 110 V Liquid Scintillation Counter


TRI-CARB 4810TR 110 V Liquid Scintillation Counter by PerkinElmer, Inc.  product image
TRI-CARB 4810TR 110 V Liquid Scintillation Counter

PerkinElmer, Inc.

The supplier does not provide quotations for this particular product through SelectScience. You can search for similar products in our product directory.



0 Scientists have reviewed this product


Write the First Review

No Reviews

The Tri-Carb 4810TR liquid scintillation counter is ready to go for basic research and CPM/direct DPM applications.

Exclusive Standard Features

  • TR-LSC (Time-resolved Liquid Scintillation Counting) for high sensitivity, low background liquid scintillation counting
  • QuantaSmart Software a robust multitasking, easy networking environment with unlimited assays in a secure multiuser environment
  • SpectraWorks 2 spectral analysis package that determines optimal counting regions, and calculates figure of merit automatically
  • Dynamic color correction ensures accuracy of tracking lower energy sample spectra over a wide quench range
  • Auxiliary Spectrum Memory stores rejected events for future analysis
  • Bi-directional sample conveyer with a capacity of up to 408 (20mL vials) or 720 (4 or 7 mL vials)
  • Operational-status LED indicator for clear assay status updates at a glance

Additional Standard Features

  • Luminescence Detection Reporting percent luminescence is flagged on printout to alert user of possible sample problems
  • Direct DPM determines single-label DPM in any cocktail without the use of quench standards
  • 15 Quick Count protocol flags with the flexibility to define unlimited assays
  • Fold-away ergonomic arm adaptable to enter data either sitting or standing
  • Built in Computer with Windows 10 Operating System

Physical data

  • Dimensions H/W/D 18.5 /40.5/32 in (47/103/81 cm)
  • Weight 477 lb (217 kg)
  • Electrical requirements 100-240 Vac 50/60 HZ 3-prong grounded plug

Typical performance data

  • Energy range: 0-2,000 keV
  • Efficiency, Normal Count Mode (minimum acceptable):
    3H (0-18.6 keV): 60%
    14C (0-156 keV): 95%
  • Observed Background, Normal Count Mode (average):
    3H (0-18.6 keV): 17 CPM
    14C (0-156 keV): 26 CPM
  • Figure of Merit (E2/B), Normal Count mode:
    3H (0-18.6 keV): 180
    14C (0-156 keV): 380

Note: The efficiencies, backgrounds, and E2/B values for the Normal Count Mode were determined using PerkinElmer sealed large vial glass standards set P.N. 6008500 verified with NIST standard activity.

Option Package

  • Replay reanalyzes sample counts without recounting samples and
  • Varisette sample cassette for intermixing vial sizes without special adapters