1 Scientist has reviewed this product
4 out of 5
Ease of use
4 out of 5
After sales service
4 out of 5
Value for money
- Christopher Bianca
Status:
Expert Reviewer 

Member since: 2013
Organization: JRFAmerica
-
- Ease of use
- 4 out of 5
- After sales service
- 4 out of 5
- Value for money
- 4 out of 5
"Very reliable instrument and durable.
Software is easy to use.
Service is great and easy to work with.
I highly recommend it.
"
Review date: 26 Aug 2015 |
LC-4000 Series UV/Visible Detectors Using the advanced Czerny- Turnermonochromator design with optimization of the optical sys tem, baseline change due to refrac tive index is minimized.
Baseline fluctuation from changes in environmental temperature is reduced with a temperature - controlled cell, baseline drift caused by temperature fluctuation in the light source is also reduced with a temperature controlled light source. The result is excellent stability with UV-4070 UV Visible UV spectra auto-scan Automatic spectral scannning can be controlled using ChromNAV, when a peak exceeds a set threshold a scan is executed automatically. 0.24 220000 200000 the noise level +/- 0.2x10 -5 AU and drift 1x10 -4 AU/h.
Both detectors offer simultaneous dual-wavelength data acquisition and spectral scanning. Us er maintenan c e c an b e d o n e without removing the module from the system. Front-panel access is used for light source replacement and automatic wavelength calibration can be carried out using a mercury lamp. Both detectors include high speed data acquisition at up to 100Hz and can be used for conventional HPLC with the standard flow cell and for UHPLC/RHPLC when used with the SP type semi-micro cell.