In-Cell Western™ Assay Quality Assessment Using Z’-Factor

5 Jun 2018

In this application note Z’-factor experiments are performed on one or more ICW assay plates containing replicate wells designated for background subtraction, negative control samples, and positive control samples. During ICW assay optimization, it is important to assess the overall quality and reliability of the assay. The Z’-factor statistic provides a way to evaluate whether or not assay conditions (reagents, protocols, instrumentation, kinetics, and other conditions not directly related to the test compounds) are optimized. Z’-factor is a dimensionless value that represents both the variability and the dynamic range between two sets of sample control data.

Odyssey DLx Infrared Imaging System

LICORbio

The Odyssey DLx Imaging System delivers accurate, reproducible, and quantitative data with near-infrared fluorescence. It supports Western blots, In-Cell Western™ Assays and other plate-based assays, cell-based assays, EMSAs, DNA and protein gels, protein arrays, and tissue section imaging. Discover more with the groundbreaking Odyssey technology cited in over 120,000 peer-reviewed publications.

(0)

Odyssey M Imaging System

LICORbio

The Odyssey M is the most versatile and sophisticated biomolecular imager on the market. With 19 available channels, industry-leading 5 µm resolution, >6 logs of dynamic range, and an onboard luminescence module, the Odyssey M gives researchers a multimodal imaging solution with unparalleled capabilities and performance. Discover more with the groundbreaking Odyssey technology cited in over 40,000 peer-reviewed publications.