- Ease of use
- After sales service
- Value for money

Related video
Description
Thermo Scientific Accela - HPLC and U-HPLC in a Single System
The Thermo Accela™ high-speed chromatographic system provides fast, efficient chromatographic separations over an expansive range of flow rates and pressures. Thermo Accela is designed to optimize performance of sub-two micron particle columns, providing seamless operation spanning conventional LC pressures from short LC columns, up to 15,000 psi for long-column separations of complex bio mixtures.
Features of the Thermo Accela include:
- The Accela Pump assures rapid and reproducible transfer of even the most complex and aggressive gradients. This quaternary pump is capable of handling pressures up to 15,000 psi with a delay volume of only 65 µL, enabling high speed chromatographic separations.
- The Accela Autosampler has a specialized diamond coated high pressure valve that can handle the rigors of constant high pressure injections while the unique sampling design enables 30 second injection cycles. This autosampler integrates isothermal injection and separation to provide superior reproducibility by eliminating all external environmental influences to the chromatography.
- The Accela PDA has been optimized for the detection of high speed chromatographic separations. The short 1 cm flow cell pathlength combined with the incredible minimized flow cell volume of 2 µL, renders this Patented LightPipe™ flowcell as the premium choice for fast separations. The extremely low level of dispersion in the LightPipe flowcell retains the excellent peak shape and chromatographic resolution from the column.
Thermo Scientific Accela, coupled with the sub-two micron particle columns, provides fast, controlled separations with high efficiency and resolution, accelerating LC and LC/MS applications.
The High Speed LC System for today’s modern laboratory. Advantages of the Thermo Accela include:
- Optimized system delay volumes for fast separations
- Flexible sample format accepts vials or plates
- Quaternary pump capable of conventional and ultra-high pressures
- Patented LightPipe technology provides maximum sensitivity and resolution
- Complete chromatography solution with Thermo mass spectrometers
Application notes and news
- Analyzing Phenolic Pollutants in Water Using U-HPLC
- Quantitation of 12 Benzodiazepines and Metabolites in Urine Using Ultrahigh Resolution LC-MS for Forensic T...
- Analysis of Triazine Herbicides in Water Using LC-MS/MS and TraceFinder Software
- Analysis of Simvastatin Tablets by High Speed LC
- Use of Small Particles in Ultra High Pressure Liquid Chromatography
- Thermo Scientific features Accela™ – HPLC and U-HPLC in one, efficient, limitless system
Reviews
MOHINEESH CHANDRA![]() Status: See all reviews Member since 2013 Organization: ALL INDIA INSTITUTE OF MEDICAL SCIENCES (AIIMS) Read more » | RATING: 4.0 ![]()
"Instrument is easy to handle and service is good. Its software is always good. Easy to handle, maintain, precise etc. I recommend this product to everyone. Being a student it’s great to learn from thermo scientific hplc." |
Kiran Oza![]() Status: See all reviews Member since 2013 Organization: Kiran Light Laboratories Pvt. Ltd Read more » | RATING: 4.7 ![]()
"Easy to use. I would recommend this to a colleague. Reliable and gives good results. Great value for money." |
Lalit Ghosh![]() Status: See all reviews Member since 2013 Organization: Ranbaxy Labs Inc Read more » | RATING: 3.0 ![]()
"This company is very reliable with good distribution. They have a good customer representative. Product quality is very good." |
Dana Bors![]() Status: See all reviews Member since 2010 Organization: Indiana University Read more » | RATING: 3.6 ![]()
"The separations are quick and efficient. The selectivity for isomers of explosives is high and both specificity and sensitivity are excellent." |
Zhaorui Zhang![]() Status: See all reviews Member since 2010 Organization: Purdue University Read more » | RATING: 4.0 ![]()
"Easy to use, straight forward operation of software and assembly of instrument. Good value for money. Normal quality, unstable low flow rate mode." |
![]() Status: See all reviews Member since 2010 Organization: SAIC-Frederick, Inc. Read more » | RATING: 2.6 ![]()
"The Accela pump that we have has too much dead volume. The user has little choice to change the function and carry over is a big issue." |
Xavier Fretté![]() Status: See all reviews Member since 2010 Organization: University of Southern Denmark Read more » | RATING: 3.0 ![]()
"The equipment has been working 3 years now. Until now we have experienced only one sequence which was interrupted during a week-end. It may sound like a minor problem, but when we have samples which are unique and that the compounds we analyze may degrade upon standing, it was very unfortunate. The machine stopped for no obvious reason, this error has not occured again, so we only have to cross fingers for it not to happen again. Furthermore the injector syringe is not airtight (it was the 1st year), which means that it is important to remember to flush the syringe before starting a sequence. Apart from these small problems, the machine is reliable and robust and we do have reproducible results. If there should be a complaint about the Thermo system, it regards the service. We also have Agilent HPLCs (1100 and 1200) though not hyphenated to MS. When we ask for service from Agilent it is usually performed within 2-3 days. With thermo it usually takes 4-7 days." |
Feedback
Request info
Write a review
Sharing your experience will help scientists like you. Achieve Reviewer Status and Win an iPad 3 (All reviews published will be entered into the next drawing on May 31st 2013).


(what's this)