If this is your profile Login now.

Mohamed Bhugun

This person does not have a profile.

  • 1 Next 
    AU680 Chemistry System
    "AU680 is a good and reliable instrument for medium sized lab but the software is not very user friendly. It's complicated for the end-user where there are too many layers to find information. Also, the ISE can't be calibrated while analyzer is running. It has to be on Standby mode. Same thing for reagent loading where it can't be loaded on the fly. So, if you run out of reagent, you have to wait for analyzer to go to standby mode which takes around 15 mins. Otherwise, the analyzer is very reliable and less troubleshooting needed. "

    Application area: Chemistry Lab

    31 May 2016
    • Ease of use3 out of 5
    • After sales service4 out of 5
    • Value for money3 out of 5

    3 out of 5

    Read more

    UniCel® DxC 800 Synchron® Clinical System
    "Good instrument, an open system which is easy to use."
    2 August 2012
    • Ease of use4 out of 5
    • After sales service4 out of 5
    • Value for money4 out of 5

    4 out of 5

    Read more

  Contact us if you have any questions about your membership or profile. Remember, membership is free!