Top List: Exclusive Advice: Choosing the Right Filter for Your MS System

13 Oct 2014


By Vivek Joshi, Ph.D., Principal Research Scientist, EMD Millipore

Membrane filtration of samples can improve the sensitivity and reproducibility of your mass spectra. These five easy ways to choose the right syringe filter for your mass spectrometry sample take into account sample volume, chemical compatibility, particle load, extractables, and analyte binding.


Choose the right size filter


The ideal filter diameter (between 4 mm and 50 mm) balances the filtration performance with the risk of extractables and analyte binding, and depends on your sample volume. While larger filters enable fast filtration using low pressure, minimizing the chance of bursting, they may have higher hold-up volume, trapping precious samples. By using larger filters, you also risk losing analyte to nonspecific binding and introducing higher levels of extractable impurities.
Use this table to choose the right filter size for your sample volume.

Sample Volume

Filter Size

Hold up Volume

Filtration Area

< 1 mL

4 mm

10 µL

0.1 cm2

1-10 mL

13 mm

25 µL

0.65 cm2

10-100 mL

25 mm

100 µL

3.6 cm2

10-100 mL

33 mm

80 µL

4.5 cm2


Table 1: Sample volume and appropriate filter size

Lots of samples?
When preparing large numbers of samples, using syringe filters or ultrafilters that filter one sample at a time can be very time consuming. Instead, try a multiwell filter plate. Filter 96 or 384 samples at once into a collection plate, and analyze directly on an LC-MS system

Choose a chemically compatible filter

If you use a filter where the membrane or the housing material is not compatible with the sample can lead to incomplete filtration, extraction of impurities into the sample, and even complete membrane disintegration. For example, some surfactants are incompatible with HDPE housings, and polyethersulfone is incompatible with acetonitrile. For mass spectrometry, even low levels of leached impurities can interfere with analysis, so refer to the website of your filter manufacturer for a complete chemical compatibility chart and extractables data.

Use prefilters for particulate-laden or viscous samples

If single membrane filters get clogged before your sample can be passed through, use a syringe filter, which contains a prefilter. The prefilter helps to trap larger particulates, protecting the final membrane from fouling and allowing higher volumes of sample to be filtered before the filter clogs. With a multilayered filter, you can filter 4-6 times more volume than a membrane-only filter

Prefilter caveat
Before you go out and try a prefilter, though, remember that prefilters are usually made of glass fibers, which can lead to higher levels of extractables or analyte binding. In such cases it might be better to filter your sample through a filter that contains an inert / low protein binding filter and a prefilter.

Choose a filter with low analyte binding

Depending on the chemical nature of your analyte, the membrane in your syringe filter might nonspecifically bind the analyte, resulting in inaccurate quantitation, especially for low concentration samples. Glass fiber filters, especially, tend to bind strongly to proteins, peptides, and oligonucleotides. Membranes vary in binding properties: nylon membranes show very high protein binding (~ 225 µg/cm2), whereas polyvinylidene fluoride (PVDF) membranes show very low protein binding (~ 15 µg/cm2). For small molecular weight analytes, hydrophilic PTFE & PVDF membranes show very low analyte binding.

Rinse filter with solvent to minimize extractable contamination

Even though signals from extractable impurities can pose problems in mass spectra, these can be reduced significantly by just washing the membrane filter with the sample or a solvent prior to filtration. In our studies of the effects of prerinsing, we found that just by rinsing the membrane with one mL of solvent, the extractables were reduced to levels undetectable by LC-UV. The ideal option is to choose the syringe filter that has low level of extractables by LC-UV and LC-MS. One such membrane is the hydrophilic PTFE syringe filter.

Ready to choose the right filter for your mass spectrometry analysis? Download the Millex® Filter Finder App!