Premier Hb9210™ - HPLC HbA1c Analysis by Trinity Biotech Plc

Manufacturer Trinity Biotech Plc
3.0
/
5.0
  |  3 reviews
The Premier Hb9210™ offers you fast, reliable & safe HbA1c diagnosis in a compact, high-capacity platform. Utilising our patented boronate affinity technology the Premier Hb9210™ offers you accurate & precise results, free from interference, in a rapid & highly scalable package. Delivering substantial workflow benefits, the platform is ideal for the most demanding laboratory & clinic environments. The instrument consists of an integrated HPLC system, a compact sample handler and... Read more
The supplier does not provide quotations for this product through SelectScience. Browse these related products instead


Premier Hb9210™ - HPLC HbA1c Analysis by Trinity Biotech Plc product image
Premier Hb9210™ - HPLC HbA1c Analysis

The supplier does not provide quotations for this particular product through SelectScience. You can search for similar products in our product directory.




Average Rating: 3.0
3 Scientists have reviewed this product

3 out of 5
Ease of use
3 out of 5
After sales service
3 out of 5
Value for money


  • Status:

    Reviewer
  • Member since: 2020

  • Organization: Nema Pathology Centre



  • Ease of use
    3 out of 5
    After sales service
    3 out of 5
    Value for money
    3 out of 5
Share this Review on LinkedinShare this Review on TwitterShare this Review on Facebook
Results are good.
Rating: 3.0

  • Application Area: HbA1C

"Need to improve after sales services. Product should have a long extended warranty but not provided by company."

Review date: 09 Feb 2021 | Premier Hb9210™ - HPLC HbA1c Analysis
  • Status:

    Reviewer

  • Member since: 2013

  • Organization: Harris County Hospital District



  • Ease of use
    2 out of 5
    After sales service
    2 out of 5
    Value for money
    2 out of 5
Share this ReviewShare this Review on TwitterShare this Review
Rating: 2.0
"We’re finding this instrument to be troublesome with calibrating and obtaining valid QC values. Calibrators and controls do not appear to be stable with the manufacture's stated life of these reagents. Technical service very limited. Do not like the fact that the instrument shuts totally down after four hours of use or has to be paused to conserve reagents."

Review date: 26 Nov 2014 | Premier Hb9210™ - HPLC HbA1c Analysis
  • Status:

    Reviewer

  • Member since: 2013

  • Organization: Harris Health System



  • Ease of use
    3 out of 5
    After sales service
    4 out of 5
    Value for money
    5 out of 5
Share this ReviewShare this Review on TwitterShare this Review
Rating: 4.0
"This instrument is very easy to use after initial calibration and controls are run. Patients can be loaded on and run started with no additional steps. Requires quite a few different reagents, but testing can be done as a stat request. Results are received very quickly."

Review date: 29 Aug 2014 | Premier Hb9210™ - HPLC HbA1c Analysis
The Premier Hb9210™ offers you fast, reliable & safe HbA1c diagnosis in a compact, high-capacity platform.

Utilising our patented boronate affinity technology the Premier Hb9210™ offers you accurate & precise results, free from interference, in a rapid & highly scalable package. Delivering substantial workflow benefits, the platform is ideal for the most demanding laboratory & clinic environments.

The instrument consists of an integrated HPLC system, a compact sample handler and the workstation with our Affinity™ control software as pictured.

Combined with a touch-screen user-interface & easy to interpret results, this advanced system allows the user to accurately determine the HbA1c results for any workload.

Unlike the charge-based separation of ion exchange HPLC, which doesn’t detect all glycated haemoglobin (Hb) species, the Premier Hb9210™ uses the glycation specific binding of boronate affinity to detect all of the glycated Hb species present.

This allows the final result to be determined based on a simple peak area fraction of two species of Hb, making result interpretation extremely straightforward as there are only two peaks on the chromatogram: one glycated (directly correlated to the HbA1c result), the other non-glycated (all other haemoglobin types).

Product Overview

Premier Hb9210™ - HPLC HbA1c Analysis by Trinity Biotech Plc product image

Premier Hb9210™ - HPLC HbA1c Analysis

Manufacturer Trinity Biotech Plc

3.0 / 5.0 | 3 reviews