CONTROL FEATURES
Delayed Start Feature Enables the autoclave to be loaded at the end of the day and be set to start automatically in the early hours of the following morning. This is so the autoclave would have already run through it’s first cycle and media or instruments are ready to work with when scientists arrive in the morning. Load Activated Sterilising
A load probe is placed into the largest load and stops the timer from starting until the load and chamber have reached set point temperature. This feature can be switched on/off for over-sensitive media.
Timed Free Steaming
The autoclave will heat up in two stages. Stage 1 will take the autoclave to around 95 degc. The chamber will then free steam for a pre-set period (specified by the operator) and then in the second stage to set point temperature. This helps for good steam penetration into the load and helps prevent overcooking of media. This can be switched on/off and is variable, pre-set into the program by the operator.
Media Stay-Warm for Media Pouring
Keeps the chamber warm so media can still be poured for some time after cycle complete.
Smart card control
Operator access to the autoclave control system by the operator by means of a smart card. The following cards are available: -