Status:
Advanced ReviewerMember since: 2019
Organization: Case Western Reserve University
Best column for protein analysis.
Application Area: Protein analysis
"This column is perfect for different protein analyses, from Insulin (small peptides) to BSA (large globular proteins). It gives sharp peaks with good reproduciblity. The carbon load of the column is just what it needed and not too much to result in tailing. It is good for academic as well as industrial setups. It definitely lasts for more than 5000 injections if maintained well. It separates impurities perfectly."
Status:
ReviewerMember since: 2016
Organization: Silliker Laboratories
Accurate Cyanocobalamin analysis by AOAC 2014.02 requires Eclipse C18 Plus column.
Application Area:Vitamin B12 analysis
"We tried to evaluate performance of the AOAC 2014.02 method in analysis of cyanocobalamin in infant formula and adult nutritionals. The method using the listed column provided a shoulder of unknown compound on cyanocobalamin peak. When we replaced the listed column by the Exclipse C18 Plus column we could find that shoulder on cyanocobalamin peak separate as a separate peak which later on was identified as adenosylcobalamin peak. Thus Eclipse C18 plus column improved the method performance remarkably and could not only estimate cyanocobalamin amount in the sample correctly but could also also analyze adenosylcobalamin."