Previously it was necessary to use standard cuvettes with a volume of at least 1 ml and additionally to dilute the sample manually. DiluCell™ from Implen allows for lower sample volume requirements and provides automatic virtual sample dilution.
Due to the unique design of DiluCell™ cuvettes, the light-path is reduced from the standard 10 mm to 0.5 mm (DC 20) or 1.0 mm (DC 10). According to the Lambert Beer Law this shortened pathlength results in an automatic sample dilution by factor 20 (DC 20) or 10 (DC 10). Automatic dilution with DiluCell™ saves time and excludes dilution errors and cross contaminations.