HPLC Columns Buying Guide

HPLC Columns image

If you are looking to invest in an HPLC column, the new SelectScience HPLC Columns Buying Guide provides important information to help you make the right decision.

Learn about key factors and application considerations when buying a new column and read real user reviews. Download the free guide »

1. Basic Overview
2. Selecting a Column: Consider your Application
3. Selecting a Column: Stationary Phase
4. Selecting a Column: Column Chemistry
5. Selecting a Column: Separation Mode
     5.1 Separations Based on Polarity
     5.2 Separations Based on Charge
     5.3 Separations Based on Size
6. Selecting a Column: Particle Properties
7. Selecting a Column: Retention Capacity
8. Selecting a Column: Column Dimensions
9. Future Trends
10. Summary

Basic Overview

An High Performance Liquid Chromatography (HPLC) column is considered to be the most important part of any instrument as this is where the separations occur. In many cases columns are at the cutting edge of separation technology, but are all basically an improved form of standard column chromatography. The main difference is that instead of a solvent being allowed to pass through a column under gravity, it is forced through under high pressure which makes separations much faster.

The higher pressure allows for the use of a smaller particle size for the column packing material. Columns with smaller particles generate sharper peaks with increased resolution due to better packing and reduced diffusion distances for analytes. This allows for faster separations versus low pressure columns.

Ultra High Performance Liquid Chromatography (UHPLC) uses an even smaller particle size than HPLC for the packing material. This has the advantages of producing increased resolution and sensitivity. UHPLC requires even higher pressures than HPLC which means systems capable of handling this increased pressure can be relatively expensive. It is worth noting that methods developed for HPLC do not readily transfer to UHPLC and vice versa.

In selecting the most appropriate column, a number of considerations must be taken into account. These are summarized in Figure 1. Further details on each consideration will be given in the coming sections.

HPLC Figure 1


Figure 1: Overview of HPLC column selection.

Selecting a Column: Consider your Application

In order to select the right column, the first thing to consider is your application. To match your application to a column there are a few simple steps to work through, the first of which is to calculate the molecular weight of your analyte. In general, compounds are divided into two groups: low molecular weight, <5000 daltons, and high molecular weight, >5000 daltons molecular weight. Assigning an analyte to one of these should be relatively easy. Secondly, you’ll need to find out if the analyte is soluble in water (polar) or an organic (non-polar) solvent. Lastly, the mode of separation can be determined to narrow column choice.

The process is summarized in Figure 2 and works as follows: if your analyte is <5000 g/mol then you should start at the low molecular weight cell, if it is water soluble move on to that cell. This results in a number of separation modes that may be used. Further details on this are given in the coming section

Molecular Weight

Analyte Solubility

Separation Mode

Low MW

(<5000 g/mol)

Water Soluble

Ion Exchange

Reversed Phase

HILIC

Organic Soluble

Normal Phase

Reversed Phase

Non-Aqueous Size Exclusion

High MW

(>5000 g/mol)

Water Soluble

Aqueous Size Exclusion

Ion Exchange

Hydrophobic Interaction

Reversed Phase

Organic Soluble

Non-Aqueous Size Exclusion

Figure 2: Determination of Separation Mode.

HPLC and UHPLC are both routinely used within analytical research from pure chromatographic studies to life science research and drug discovery. When used in drug discovery, for example, it has been shown to significantly improve the success of the process. It can prevent poor drug candidates from progressing through the discovery process by monitoring factors such as metabolic stability and toxic metabolite production. The technique has also been successfully applied to the separation and analysis of challenging samples such as soil, environmental waters and explosives which are of importance to environmental and forensic laboratories. The usefulness of HPLC and UHPLC can be further enhanced by linking to other systems such as mass spectrometers.

There are many different columns on the market that have been designed for specific applications, such as: Enantiomers; Vitamins; Pesticides; Carbohydrates. It is worth investigating if there are already columns available that have been optimized for your application.

Selecting a Column: Stationary Phase

Once the application has been identified, choosing the best column for your application is based on two considerations. The first and most important of these is the type of stationary phase. The factors to be considered in this regards are the column chemistry, the separation mode, the particle size and the retention capacity.

Selecting a Column: Column Chemistry

In general columns are filled with porous particles coated with a material that interacts with the injected sample. HPLC uses what is called a true stationary phase: column “chemistries” are bonded tightly to the packing material and do not bleed off.

There are various ways to categorize the types of stationary phase available but the most widely accepted of these is the USP “L” system. This is a list of over 60 column classifications ordered according to the type of bonded phase (C18, C8…), packing material and particle size. Figure 3 shows some of the most common types.

USP

Stationary Phase

Details

L1

C18 - Octadecyl silane

Octadecyl silane bonded to porous silica, 1.5-10 µm

L7

C8 - Octyl silane

Octyl silane bonded to totally porous silica, 1.5-10 µm

L8

NH2 - Aminopropyl silane

Aminopropyl-silane bonded to totally porous silica, 3-10 µm

L10

CN - Nitrile group

Nitrile groups bonded to porous silica, 3-10 µm

L11

Phenyl

Phenyl groups bonded to porous silica, 1.5-10 µm

L43

PFP - Pentafluorophenyl

PFP groups bonded to porous silica, 5-10 µm

Figure 3: Common USP ‘L’ Classifications.

Selecting a Column: Separation Mode

The column chemistry selected is also dependent upon the separation mode which is going to be used. In general, three primary characteristics of chemical compounds can be used to create HPLC separations. They are:

1.) Polarity
2.) Electrical Charge
3.) Molecular Size

1.) Separations Based on Polarity

There are two primary separation modes that are commonly used to exploit the polarity of a chemical compound. These are Normal Phase and Reversed Phase. Both are dependent on the relative polarity of the solvent used and the stationary phase. Reversed phase HPLC is the most commonly used form. In recent years, Hydrophilic Interaction Chromatography (HILIC) and Hydrophobic Interaction Chromatography (HIC) have become more widely used, in particular for the separation of polar compounds and large biomolecules, respectively.

Normal Phase HPLC.: Normal-phase chromatography, or NP, is the classic form of liquid chromatography using polar stationary phases. and non-polar mobile phases. The analyte is retained by the interaction of its polar functional groups with the polar groups on the surface of the packing. Analytes elute from the column starting with the least polar compound followed by other compounds in order of their increasing polarity. Normal-phase chromatography is useful in the separation of analytes with low to intermediate polarity and high solubility in low-polarity solvents. Water-soluble analytes are usually not good candidates for normal-phase chromatography.

Reversed Phase HPLC: In reversed phase chromatography, or RP, the stationary phase is modified to make it non-polar by attaching long hydrocarbons to its surface. Typically these hydrocarbons will be hydrophobic alkyl chains of 4, 8 or 18 carbon atoms in length which interact with the analyte. Manufacturers commonly name their columns depending on the number of carbon atoms i.e. C4, C8, C18. Of these chain lengths C4 is generally used for separating proteins whilst C8 and C18 chemistries are used for peptides or small molecules. In general reversed phase columns work well for water-soluble hydrophobic compounds.

The addition of these alkyl chains means that polar molecules in the mixture will spend most of their time moving with the solvent. Non-polar compounds will form attachments with the hydrocarbon groups and therefore spend less time in the solvent; this slows them through the column. So in contrast with the normal phase, column polar molecules will travel through more quickly in reversed phase columns. RP chromatography is useful for the separation of compounds having high to intermediate polarity.

Hydrophilic Interaction Chromatography (HILIC): HILIC is a relatively recently adopted HPLC phase, which is used for the retention of water soluble (polar) compounds. Liquid-liquid partitioning between the polar mobile phase and the water layer at the stationary phase surface is considered to be the main mechanism of action. In this type, polar stationary (e.g. silica and amine) and mobile (e.g. water and methanol) phases are used and an increase in mobile phase polarity results in a decrease in analyte retention. This makes it very useful in separating highly polar compounds, which would be unretained by reversed phase chromatography. Amongst the most typical analytes for HILIC are amino acids, polar pharmaceuticals, nucleobases, nucleosides, alkaloids and carbohydrates.

Hydrophobic Interaction Chromatography (HIC): HIC, the most recent addition to the different modes of chromatography, is a type of reversed-phase chromatography that is used to separate large biomolecules, such as proteins. It is usually desirable to maintain these molecules intact in an aqueous solution, avoiding contact with organic solvents or surfaces that might denature them. The promotion of the hydrophobic effect (by addition of lyotropic salts) drives the adsorption of hydrophobic areas on a protein to the hydrophobic areas on the solid support. This is thermodynamically favorable in that it reduces the number and volume of individual hydrophobic cavities. Reducing hydrophobic interaction, by decreasing the concentration of lyotropic salts, results in de-sorption from the solid support.

HIC is unique in that proteins bind at high salt concentration and elutes proteins at low salt concentration. This is manifested in a reverse salt gradient which is an immediate indication that HIC is being employed.



2.) Separations Based on Charge

Ion-Exchange Chromatography (IEC or IEX) separates analytes based on differences in their surface charge. It is typically used for the analysis of biological samples such as proteins, peptides, amino acids, nucleic acids and glycoproteins. The basic principle governing IEC separations is that analytes with an electrical charge are attracted to ligands on the stationary-phase particles, which bear the opposite charge. The mobile phase typically consists of a pair of aqueous buffers, with the one buffer containing a low concentration of salt and the other buffer containing much more salt.

The stationary phases for ion-exchange separations are characterized by the nature and strength of the acidic or basic functions on their surfaces and the types of ions that they attract and retain. Two types of ion-exchange chromatography exist. Cation exchange is used to retain and separate positively charged ions on a negative surface. Conversely, anion exchange is used to retain and separate negatively charged ions on a positive surface.

Anion and cation exchangers are further classified as strong or weak, depending on how much the ionization state of the functional groups vary with pH. A strong ion exchanger has the same charge density on its surface over a broad pH range, whereas the charge density of a weak ion exchanger changes with pH.

The choice of whether to use a strong verses weak cation or anion exchanger depends on the charge of the biomolecule of interest. For example, a weakly acidic molecule, which requires very low or very high pH for ionization, will require the use of a strong ion exchanger because it functions at extremes of pH. You will also need to consider the stability of your molecule. If the substance is stable to extremes of pH (eg. amino acids) strongly acidic or basic ion exchangers can be used. If substance is labile (eg. proteins) a weaker exchanger of low charge density is advisable because the molecules can be eluted by more gentle conditions of pH and ionic strength. When choosing an appropriate column for IEC, it is advantageous to have knowledge of the PI of your molecule of interest.



3.) Separations Based on Size

Size-Exclusion HPLC (SEC) separates analytes on the basis of a combination of their hydrodynamic size, diffusion coefficient, and surface properties. It is usually applied to large molecules or macromolecular complexes such as proteins and industrial polymers.

In SEC, the stationary phase is composed of inert particles packed into a dense three-dimensional matrix within a glass or steel column. The stationary-phase particles have small pores and/or channels which will only allow species below a certain size to enter. Smaller molecules will penetrate more of the pores on their passage through the bed and will take longer to elute. Larger molecules can only penetrate pores above a certain size so they will spend less time in the bed. The biggest molecules may be totally excluded from pores and pass only between the particles, eluting very quickly in a small volume. The mobile phase can be pure water, an aqueous buffer, an organic solvent, or a mixture of these.

There are two basic types of size exclusion chromatography. One is gel permeation chromatography (GPC), which uses a hydrophobic column packing material and a non-aqueous mobile phase (organic solvent) to measure the molecular weight distribution of synthetic polymers. The other is gel filtration chromatography (GFC), which uses a hydrophilic packing material and an aqueous mobile phase to separate, fractionate, or measure the molecular weight distribution of molecules soluble in water, such as polysaccharides and proteins.

There are many different commercially available SEC columns which have different packing material pore sizes (different exclusion limits). When choosing a column you should compare the calibration curves and select a column that is best suited to the range of molecule weights to be measured. If your sample is distributed over a wide range of molecular weights you may want to consider using several columns connected in a series. Many SEC columns have limitations on the eluents that can be used. Using an eluent that is not specifically permitted could irreversibly expand the packing material or cause physical deterioration. You should therefore investigate which solvents readily dissolve the sample in advance and then choose a column compatible with one of those solvents.

Selecting a Column: Particle Properties

The next consideration is the size and shape of the particles which make up the column packing material. The majority of new HPLC methods are performed on spherical shaped or spheroidal (almost spherical) particles. Spherical particles provide higher efficiency, better column stability and lower back-pressures compared to irregularly shaped particles.

If all other factors are constant, the smaller the particles are the more efficient the separation. The major problem with a decrease in particle size, from 10-, 7-, 5-, and 3-micron diameters, is that the backpressure increases exponentially. This roughly means a column using 3-micron particles is about twice as efficient as a 5-micron column, but pressures are three times as high.

Although separation efficiencies can be increased by reducing particle size even more, to below 2 microns, the hardware required becomes more expensive as it needs to handle the extremely high pressures. UHPLC systems have been designed to overcome these limitations and UHPLC columns can deliver shorter run times, cleaner separations, sharper/taller peaks, and improved detection limits.

Some applications also benefit from end-capped HPLC columns. A reversed-phase HPLC column that is end-capped has gone through a secondary bonding step to cover unreacted silanols on the silica surface. End-capped packing materials can eliminate unpredictable secondary interactions.

Selecting a Column: Retention Capacity

The retention capacity is the time samples spend in the column. It is of importance to both the quality and speed of separations. A column which takes a long time to elute will not be of use to a high-throughput application.

The retention capacity is influenced by surface area and carbon load (the percentage of carbon in the packing material). The pore size and volume of the packing material is also important in determining retention capacity. This is because surface area is inversely proportional to pore size. A larger pore size results in lower retention. The particle pore size is measured in angstroms and generally ranges between 100-1000 angstroms. 300 angstroms is the most popular pore size for proteins and peptides and 100 angstroms is the most common for small molecules.

Depending on your requirements HPLC columns can be quite expensive and you will therefore want to ensure they last as long as possible. Guard columns can be used to protect your analytical column and prolong their lifetime. HPLC guard columns (or cartridges) are installed in front of an analytical column and protect it from strongly retained impurities. An alternative to using a guard system is the use of an in-line filter. Filters are a simple cost-effective means to providing a quick efficient clean up for occasional 'dirty' samples.

Selecting a Column: Column Dimensions

Once a packing material has been selected, the physical dimensions of the HPLC column hardware should also be optimized for the desired separation. Larger columns are useful in the scale up process, but smaller columns generally offer greater sensitivity and are therefore more useful in analytical applications.

The column length is determined by how many, and what type, of compounds you have to separate. Long columns provide better resolution and sensitivity but require longer retention times and higher pressure. It is best to choose the shortest column possible for your application without affecting resolution.

The internal diameter (ID) of an HPLC column influences both detection sensitivity and separation selectivity (resolution). The most commonly used ID for HPLC columns is 4.6mm. Small column diameters provide higher sensitivity than larger column diameters for the same injected mass because the concentration of the analyte in the mobile phase is greater. Smaller diameter columns also use less mobile phase per analysis because a slower flow rate is required to achieve the same linear velocity through the column. The major disadvantage associated with smaller diameter columns is that the sample loading capacity is reduced.

In summary:

  • Larger ID columns (over 10 mm) are mainly used to purify large scale amounts of material because of their relatively large loading capacity.
  • Analytical scale columns (4.6 mm) are the most common type. They are most often used in quantitative analysis.
  • Narrow-bore columns (1–2 mm) are used for applications where more sensitivity is required.
  • Capillary columns (under 0.3 mm) are used almost exclusively with alternative types of detection such as mass spectrometry.

Choosing the physical dimensions of your column will therefore depend on your application and throughput requirements.

Future trends

The most recent development in HPLC is the use of fused-core column technology. It is claimed to provide the performance of sub-2-micron particles, higher separating efficiencies similar to UHPLC, but at normal pressures. This technology uses a solid silica particle covered with a layer of porous silica, which is then infused with the bonded phase. A big advantage of fused-core columns over UHPLC is that while methods developed on UHPLC or conventional HPLC cannot be transferred unless the two labs have the same instrument, with fused-core particle columns different labs only need to have the same type of column.

With the advent of a second generation of silica monolith columns, liquid chromatographers now have an alternative to fused-core packings for high-throughput, high-efficiency separations. The more favorable correlation of column back pressure and separation efficiency, shown by fused-core particles as compared to fully porous particles, can now also be achieved with a new generation of silica-based monolithic HPLC columns. While monolith columns have recently emerged as a viable alternative for liquid phase separations they are still being developed for certain applications.

In the next few years, as HILIC and HIC continue to become more widely used, we are likely to see an increase in the number of phases that are available for these techniques. HILIC is especially favored by mass spectroscopists since ionization efficiency is often enhanced in organic solvents and the presence of low or no buffer salt compared to reversed-phase chromatography.

Traditionally, separations are performed on straight phases where one main type of chromatographic interaction governs the separation. To achieve different selectivity, column researchers are now synthesizing phases that have two or three different functionalities. Mixed-mode phases are becoming more widely accepted and may play a future role in enabling scientists to achieve separations that were not previously possible.

Although silica gel and silica hybrids with chemically bonded phases still dominate the applications of HPLC, newer packing materials continue to be developed. Whether the particle size reduction trend will continue is a matter of speculation. Smaller particles will require even higher pressure and place additional constraints on the packing materials, column hardware and instrument consumables. It remains to be seen whether fused-core column technology and second generation monoliths will take over from traditional packing materials.

Summary

There are many different types of HPLC column on the market and finding the correct one for your application may seem daunting but, by applying a few simple considerations, selecting the correct column can be made an easier task.

HPLC columns will continue to develop and new technologies will emerge. Visit the SelectScience product library to find out about the latest HPLC columns from leading manufactures and read user reviews. Use the SelectScience application note library to keep up to date with the latest HPLC methods.

Editor's picks

Kerry Parker

Editor

Chromolith (Merck Millipore)

Product image

5 out of 5

“One of our best performance columns in our laboratory. Good separation, long life and reproducible results.”
Ronny Dayag, Tnuva




Read more

Kinetex® 5 µm Core-Shell UHPLC Column (Phenomenex Inc)

Product image

4 out of 5

"The retention times are shorter compared to a conventional C-18 due to the new kind of pre-column Phenomenex use.”
Martin Leitenberger, Universität Hohenheim

Read more






ACQUITY UPLC CHS and XSelect Columns (Waters)

Product image

5 out of 5

"Water's CSH columns offer a significant advantage over other types of columns. It offers superior peak shape for nearly all basic compounds, increased loading capacity and rapid column equilibration after changing mobile-phase pH."
Tom Phillips, State Chemist Section, Maryland Department of Agriculture

Read more

ZORBAX StableBond (Agilent Technologies)

Product image

5 out of 5

“We had a very positive experience with this column! It gives slim peaks, good selectivity and is very stable. Ideal for HPLC and UPLC.”
Veronika Kaindl , Hexal AG

Read more



Accucore™ XL C18 HPLC Columns, 4µm particle size (Thermo Fisher Scientific)

Product image

4 out of 5

"Accucore XL HPLC columns allow users of conventional HPLC systems to change their existing methods or develop new methods with shorter run times, higher efficiency and improved resolution, within the standard pressure limits of their equipment."
Pantelis Kapetis, Metrolab SA

Read more

Ascentis® Express C18 HPLC Column (Sigma-Aldrich Corporation)

Product image

4 out of 5

"This is a coreshell column (as opposed to a fully porous). In our method for the analysis of anthocyanidins in fruit, it has the same resolution as a sub 2 µm fully porous column."
Steven Lloyd, USDA-ARS



Read more



To help you choose the correct system, use the SelectScience product and supplier directory for an overview of systems from leading manufacturers and read user reviews from other SelectScience members.

Buying guides

Advertisement

Join 200,000 Scientists Online

Become a member for FREE access to:

  • 250,000 full product and supplier listings
  • 10,000 user reviews and comments
  • Daily news and videos
  • 2,500 application notes and videos
  • 2,000 of the best jobs in Science

Don't miss out!

Join The Community Today